Incubator Bayi YP-100

No Comments on Incubator Bayi YP-100

Incubator Bayi YP-100

Incubator Bayi YP-100


Onemed


  • Other specification
    • Infant mattress width 31cm
    • Infant mattress length 63cm
    • Noise level within hood environment <60dB (A)
      (Base on surroundung environment noise below 45dB)
    • Carbon diaxide(Co2) level within the hood less than 0.5% when a mixture of Co2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress.
    • Air velocity over mattress  less than 0.35m/s
  • Reset Key
    • When the following alarm occurs, press RESET/SILENT key to silence the alarm for 5 minutes, the alarm will come back automatically after 5 minutes. If alarm condition no longer exists, the alarm will be canceled (except overheat); or press the RESET/SILENT key twice to cancel the alarm, the equipment will come back to set state
    • Temperature deviation alarm
    • Sensor alarm
    • Motor alarm
    • Over temperature

  • Temperature control range and correlative
  • Air mode control 25-37°C
  • Temperature rise time (typical @22°C ambient) < 45minutes
  • Temperature variability < 0.5°C
  • Temperature uniformity (level mattress) <0.8°C
  • Alarm

    • Power alarm power failure
    • Motor alarm  motor failure
    • Motor run speed too low
    • Air flow stop
    • Sensor alarm  Air temperature sensor failure
    • Isolate overheating sensor failure
    • Air flow sensor failed
    • Temperature deviation alarm High/low air temperature
    • Over temperature alarm Air temperature control mode
    • Activated if incubator temperature not exceeds 38°C
    • E² ROM failed  internal malfunction
  • Environment temperature (not to use in the environment exceed specified)
    • Operating range  +20-+30°C
    • Storage range -40-+70°C
  • Environment humidity
    • Operating range 30%RH-75%RH non-condensing
    • Storage range  < 80% RH non-condensing
  • Air Velocity Ambient Air Velocity  < 0.3 m/s
  • Atmospheric Pressure

    • Shipment & store atmospheric pressure range 500hPa-1060hPa
    • Operating atmospheric pressure range 700hPa-1060hPa
  • Power requirements
    • AC220V-230/AC110V-120V, 50/60Hz
  • Power requirements
    • 680VA maximum
  • Power outlet
    • AC220V-230V, 50/60Hz,Current 2A max
    • AC110V-AC120V,50/60Hz,Current 2Amax
  • Physical Dimensions
    • Height from mattress to floor 90cm
    • Width 100cm
    • Depth 150cm
    • Incubator main body weight > 60kg
    • Stand weight > 10kg
    • I.V Pole max. Load 20N
    • Hood max. Load 4kg
    • Infant mattress max. Load 10kg