Incubator Bayi YP-930

No Comments on Incubator Bayi YP-930

Incubator Bayi YP-930

Incubator Bayi YP-930


Onemed


  • OPERATIONAL CONDITION
    • Environment temp 20°C~30°C
    • Environment relative humidity 30%~75%
    • Environment air velocity of flow <0.3m/s
    • Air temp control range 25°C~37°C (temperature override mode to = 39°C)
    • Skin temp control range 34°C~37°C (temperature override mode to = 38°C)
    • Skin temp sensor accuracy <0.3°C
    • Temperature fluctuation <0.5°C
    • Mattress temperature uniformity <0.8°C (level mattress)
    • Temperature rise time <30min
    • Noise level within hood environment <55dB
    • Mattress tilt tredelenburg / reverse
    • Tredelenburg 0°~5° continuous variable
    • Mattress tray dimension 630x350mm
    • Mattress tray max.load 10kg
    • I.V pole max.load 2kg

  • OPERATIONAL CONDITION
    • Environment temp 20°C~30°C
    • Environment relative humidity 30%~75%
    • Environment air velocity of flow <0.3m/s
    • Air temp control range 25°C~37°C (temperature override mode to = 39°C)
    • Skin temp control range 34°C~37°C (temperature override mode to = 38°C)
    • Skin temp sensor accuracy <0.3°C
    • Temperature fluctuation <0.5°C
    • Mattress temperature uniformity <0.8°C (level mattress)
    • Temperature rise time <30min
    • Noise level within hood environment <55dB
    • Mattress tilt tredelenburg / reverse
    • Tredelenburg 0°~5° continuous variable
    • Mattress tray dimension 630x350mm
    • Mattress tray max.load 10kg
    • I.V pole max.load 2kg
    • Low Temp Deviation Alarm
    • Air mode -3.0°C
    • Skin mode  -1.0°C
    • High Temp Deviation Alarm
    • Air mode <38°C (temperature set point <37°C) or <40°C (temperature set point >37°C); skin mode <40°C,
    • High Temp Cut Out Alarm
    • Air mode
      • <38°C (temperature set point <37°C) or
      • <40°C (temperature set point >37°C)
      • skin mode <40°C,
    • Sensor Alarm
      • Activated if sensor does not connected, short circuit, open circuit, or does not place in the right position
    • Motor Alarm
      • Activated if motor stop, or run speed <1000rpm, air flow jam
      • Power Failure Alarm
      • Activated if power break down or power cord does not connect when opening the unit
    • System Alarm
      • Activated if internal E²ROM failed
  • TECHNICAL DATA
    • Power supply requirement AC220V, 50Hz
    • Power requirement < 600VA
  • Mattress tray dimension 630x350mm