ECG CardioCare 2000 (Bionet)

No Comments on ECG CardioCare 2000 (Bionet)

ECG CardioCare 2000 (Bionet)

ECG CardioCare 2000


Bionet


  • Over 130 kinds of interpretation results based on advanced Minnesota code
  • High-Quality 12ch interpretive ECG at a 3ch price
  • Free S/W upgrade via internet

  • ECG leads 12 lead, simultaneous ECG and acquisition
  • Recording channel : 3, 6, 12 channel and 1 channel(60 sec)
  • Sensitivity  : 5, 10, 20, auto(I~aVF:10, V1~V6:5)mm/mV
  • Printing Speed : 12.5, 25, 50 mm/S
  • Sampling rate : 500 samples/sec
  • Filters
    • AC(50/60 Hz, -20 dB or better)
    • Muscle (25~35 Hz, -3 dB or better)
    • Base line drift(0.1 Hz, -3 dB or better)
    • Low pass filter : off, 40Hz, 100Hz, 150Hz
  • Display 2 x 16 character LCD display
  • LCD display Sensitivity, speed, filter status, HR, printing form, rhythm lead
  • LED indicator Signal quality, power source, battery
  • Keyboard Membrane keyboard, Alphanumeric and symbol available
  • Patient data ID, name, age, sex, height, weight
  • Basic measurement Heart rate, PR int, QRS dur, QT/QTc, P-R-T axis
  • Recorder Thermal print head, Thermal roll paper
    • Report papers
      • Width  A4 210mm or 8.5”
      • Length A4 300mm or 11”
    • Effective writing
      • width 204 mm
      • Resolution
        • Vertical 8 dots/mm
        • Horizontal 16 dots/mm
  • Electrical
    • Internal noise 20 µV (p-p) max
    • Input circuit Floating input
    • Isolated and defribrillation protected
    • Input impedance > 10M W
    • Input voltage range > + 5 mV
    • Common Mode Rejection > + 100 dB
    • DC offset voltage > + 300 mV
    • Time constant 3.2 sec
    • Patient leakage current < 10 µA
    • Frequency response 0.05~150 Hz
  • Signal quality control : Disconnected lead detection
  • Power
    • Power supply AC or built-in Battery (option)
    • Power requirement 100-240 Vac, 50/60 Hz, 1.0-0.5 A
    • Power consumption 60W max
  • Battery capacity 1 hour of normal use (approximately 100 automatic ECG printouts)
  • Communication PC connection with RS-232 interface and LAN
  • Safety Class I, type BF
  • Conformity CE, CSA, FDA, KFDA, SFDA, CCC
  • Environment
    • Operating humidity 30 ~ 85 %
    • Operating Temperature 10° ~ 40°C
    • Atmospheric pressure 70 ~ 106 KPa
  • Standard accessory Power cord 1 EA, Patient cable 1 EA, Limb 1 set (4 EA), Chest ball 1 set (6 EA), Chart paper 1 EA, Operation manual 1 EA
  • Options Rechargeable battery, Cart, Hanger, Bag, PC S/W
  • Dimension :
    • 296 x 305.5 x 92.5 mm,
    • Approx 2.98 kg