Brosur Zeus 400
  • Use in, General surgery, Dermatology, Endoscopy, Vascular surgery, Gynecology, Heart/Thoraric Surgery, ORL/ENT, Pediatric Surgery, Orthopedics, Urology, Oral Surgery, T.U.R
  • External power generator, When the user specifies the power requirement, ZEUS is designed to provide consistent and reliable power. Our power source setting will benefit the user clinical user requirements and flexibility in operation.
  • ZEUS Smooth cut, User can have smooth cut with minimum power needed and can choose from 4 different modes of monopolar cut setting.
  • Safe with feedback system, ZEUS has special feedback system for patients and users safety
  • ZEUS Coagulation, Users can select contact or spray type coagulation which needed
  • Easy Autohand switch in bipolar mode, for easy operation, it can control with hand and foot, both ways.
  • Pure Cut
    – Max Output power : 400W at 300?
    – Carrier Freq : 400 KHz
    – Duty Rate : 100%
    – Repeat Freq : Continuous
  • Blend 1
    – Max Output power : 250W at 300?
    – Carrier Freq : 400 KHz
    – Duty Rate : 80%
    – Repeat Freq : 33KHz
  • Blend 2
    – Max Output power : 200W at 300?
    – Carrier Freq : 400 KHz
    – Duty Rate : 60%
    – Repeat Freq : 33KHz
  • Blend 3
    – Max Output power : 150W at 300?
    – Carrier Freq : 400 KHz
    – Duty Rate : 50%
    – Repeat Freq : 33KHz
  • Contact Coagulation
    – Max Output power : 120W at 300?
    – Carrier Freq : 33KHz – 153 KHz
    – Duty Rate : 100%
    – Repeat Freq : Pulse Modulated
  • Spray Coagulation
    – Max Output power : 100W at 300?
    – Carrier Freq : 400 KHz
    – Duty Rate : 8%
    – Repeat Freq : 33KHz
  • Bipolar Coagulation
    – Max Output power : 99W at 100?
    – Carrier Freq : 500 KHz
    – Duty Rate : 100%
    – Repeat Freq : Continous
  • Main Frequency : 400, 500 KHz
  • Repetion Rate : 33 KHz
  • High Frequency leakage Current : < 150mA
  • LowFrequency leakage Current
    – Patient leads to ground : < 0.01 mA
    – Chassis to Ground : < 0.1 mA
  • Power source : 230V,50/60 Hz
  • Power Consumption : 1200 VA
  • Dimension (WxDxH mm) : ± 330 x ± 455 x ± 150 mm
  • Dimension (WxDxH mm) : ± 330 x ± 455 x ± 150 mm
  • Weight : About 12.5 kg